Pipette Range Dilution Planner

Keep each step in the range your pipette can hold.

Presets:
Formula: C₁ × V₁ = C₂ × V₂
V₁ = (C₂ × V₂) / C₁
Stock Source
Calculating
Target Conc
Final Volume
Fluid color:
Dilution Factor (DF)
5.00×
1 : 5 dilution
Solvent (Buffer / Water)
40.00 mL
V₂ - V₁ volume

Pipette Range Dilution Planner: Live Flask Simulation

Live color sync
Transfer 10 mL aliquot
Stock Solution (C₁)
75%50%25%
10 mM
Concentration (C₁)
Aliquot to draw: 10 mL
Final Dilution (C₂)
CALIBRATION MARK (V₂)
2 mM
Concentration (C₂)
Total Volume: 50 mL
Color depth (Beer-Lambert) Relative Intensity: 20% (Absorbance Factor: 0.200)

Step-by-Step Bench Protocol

1

Pipette Stock Aliquot

Draw 10 mL of stock (10 mM) with a pipette.

2

Add Diluent / Buffer

Add 40 mL of water or buffer to reach the final mark.

3

Invert and Mix

Invert the tube to mix. You now have 50 mL at 2 mM (a 5× dilution).

How It Works

This planner works in 4 steps.

  1. List the pipettes

    Select or type each pipette's minimum and maximum volume.

  2. Enter the dilution

    Type the stock, target, and final volume, or a multi-step scheme.

  3. Assign pipettes

    The planner matches every transfer to the pipette that places it highest in range.

  4. Read the adjusted plan

    The planner returns pipette per step and suggests a scale-up or predilution for out-of-range steps.

The Formula

The formula checks each transfer against each pipette's range and picks the pipette with the highest fraction of nominal.

Range check V_min ≤ V_transfer ≤ V_max
Fraction of nominal f = V_transfer / V_max
Pick Pipette with the largest f within range
V_transfer:
Volume to move
V_min / V_max:
Pipette range limits
f:
Position in range (1.0 = nominal)
Worked example

A 15 µL transfer fits a P20 (75% of nominal) and a P200 (7.5% of nominal, below the 20 µL minimum). The planner assigns the P20.

Conversion Table

The conversion table lists typical ranges for 6 single-channel pipette sizes and the upper band where they perform best.

PipetteMinimumMaximum10% of nominalPreferred band
P20.2 µL2 µL0.2 µL0.7 – 2 µL
P101 µL10 µL1 µL3.5 – 10 µL
P202 µL20 µL2 µL7 – 20 µL
P20020 µL200 µL20 µL70 – 200 µL
P1000100 µL1,000 µL100 µL350 – 1,000 µL
P5000500 µL5,000 µL500 µL1,750 – 5,000 µL

Ranges are typical. Use the manufacturer's stated range for each model.

What Is the Pipette Range Dilution Planner?

The Pipette Range Dilution Planner is a tool that designs a dilution so every transfer falls within the specified range of the pipettes on hand. The user enters each pipette's minimum and maximum volume, and the planner assigns a pipette to every step.

Air-displacement pipettes are least accurate at the low end of their range. ISO 8655-2 sets tighter error limits at nominal volume than at 10% of nominal. The planner prefers transfers in the upper part of a pipette's range and splits or restructures any step that falls outside every available range.

Common Uses

There are 5 main uses of this planner.

  • Protocol design

    Developers write protocols that fit common pipette sets.

  • Teaching labs

    Instructors design dilutions for P20, P200, and P1000 only.

  • Field kits

    Field teams plan with the 1 or 2 pipettes they carry.

  • Accuracy improvement

    Analysts move transfers away from the bottom of a range.

  • SOP writing

    QA teams name the pipette for every SOP step.